Parameters:
There are two digital channels, so n can be either 1 or 2. For the trigger source,
this command expects an ASCII string. The trigger source can be “Stage”, “PSI”,
“Temp”, etc.
Response:
None.
Example:
-> DIGITAL1TRIGGERSOURCE=Stage
5
Command sent to XL™.
rs
Response from XL™.
Command
DIGITALnTRIGGERTYPE?
Description:
Read the trigger type of the n
th
digital channel. The trigger type determines the
way in which the value of the trigger source is compared to the trigger point.
Parameters:
There are two digital channels, so n can be either 1 or 2.
Response:
An ASCII string is returned containing the trigger type. The trigger type can be
“Always”, “> Set Pt”, or “< Set Pt”.
Example:
-> DIGITAL1TRIGGERTYPE?
5
Command sent to the XL™.
Digital 1 Trigger Type = Always
rs Response from the XL™.
Command
DIGITALnTRIGGERTYPE=
Description:
Write the trigger type of the n
th
digital channel. The trigger type determines the
way in which the value of the trigger source is compared to the trigger point.
Parameters:
There are two digital channels, so n can be either 1 or 2. For the trigger type, this
command expects an ASCII string. The trigger type can be “Always”, “> Set Pt”,
or “< Set Pt”.
Response:
None.
Example:
-> DIGITAL1TRIGGERTYPE=> Set Pt
5
Command sent to the XL™.
rs
Response from the XL™.
Command
DIGITALnTRIGGERPOINT?
Description:
Read the trigger point of the n
th
digital channel. The trigger point is compared to
the value of the trigger source, and if the comparison meets the condition specified
H-5223
Remote Operation (Command Mode) C-23