Parameters:

There are two digital channels, so n can be either 1 or 2. The command accepts an
integer number for the pulse length.

Response:

None.

Example:

-> DIGITAL1PULSELENGTH=150

5

Command sent to the XL™.

rs

Response from the XL™.

Command

COUNTERSLOPE?

Description:

Read the counter slope. This value scales raw counts to the value given to the user.

Parameters:

None.

Response:

The response is an ASCII text string representing the slope value. The response
will be followed by a carriage return, line feed.

Example:

-> COUNTERSLOPE?

5

Command sent to the XL™

Counter Slope = 0.01

rs Response sent to the host system.

Command:

COUNTERSLOPE=XXXXX

Description:

Set the counter slope. This value scales raw counts to the value given to the user.

Parameters:

This command accepts an ASCII string representing the new slope value.

Response:

None.

Example:

-> COUNTERSLOPE=1.0

5

Command sent to the XL™.

rs

Response sent to the host system.

Command:

RESETCOUNTERS

Description:

Reset both total and summation counters to zero.

Parameter:

None.

Response:

None.

Example:

-> RESETCOUNTERS

5

Command sent to the XL™

rs

Response sent to the host system.

H-5223

Remote Operation (Command Mode) C-25