Other programming, 11 burnout – Omega RD200 User Manual
Page 79
-75-
11. OTHER PROGRAMMING
11.11 Burnout
The channel where the temperature range is programmed becomes effective. Press and
simultaneously in the operation screen for 3 seconds or more to display the [Burnout] programming screen. This
programming is necessary for each channel.
1
Burnout
• If a sensor (thermocouple or resistance
thermometer) is disconnected, trace printing
overshoots maximum or minimum limit.
• The default parameter of burnout is “disable”
() in all channels.
Burnout can be selected for each channel out of 3
types shown below.
(Non)
: Burnout disable
(UP burn)
: Up scale burnout
(down burn)
:
Down scale burnout
2
Programmed Flow Chart
Operation screen
[Check]
( + )
3 sec. or more
[To be ready for programming]
[Specifying channel]
[Selecting burnout]
Press several times.
Temporarily stored
Programming other channels?
+
Stored
A parallel connection of a
thermocouple with other instrument
causes a problem. Select () if
your recorder uses the thermocouple
being connected with other
instrument.
Even if the burnout is programmed to
a channel where the voltage range
(No. 01 to 07) is programmed, this
programming becomes ineffective
(burnout disable).
For the burnout operation (display and
printing), read Section 10.2.
Press or to change
channels. The programmed burnout of
other channels can be checked.
Store the [temporarily stored]
parameters into memory.
(Programming change mark is
printed.)
Channel
Cursor
YES
NO
Cursor
Cursor
Burnout selection menu
“” programming on
parallel operation
SHIFT
%
DISP
%
SET
END
2
TAG
/
ENTRY
SHIFT
SET
END
/
%
SHIFT
Storing
Reference 3
Other channels check
Reference 2
Remarks 1
Voltage range is ineffective
Remarks 2
Burnout operation
Reference 1
Ref. 2
Ref. 3