Adjustment, 3 adjustment of printing position – Omega RD200 User Manual
Page 112
-108-
18. ADJUSTMENT
18.3 Adjustment of Printing Position
This is the adjustment for the printing position of the cartridge pen for use in trace printing. It is recommended to
adjust it once a year for maintaining the printing accuracy.
■
Adjustment Flow Chart
Operation screen
[Engineering mode selection]
( + )
3 sec. or more
[Selecting adjustment]
→
[Selecting printing position]
→
[Selecting pen No.]
To match the printing position to “Zero”, press few times.
(Computation of zero
correction data)
To match the printing position to “span”, press few times.
(Computation of span
correction data)
Temporarily storing
Programming other channel?
+
Stored
Pressing once moves 0.1 mm
to right. Pressing once moves
0.1 mm to left. Press when the
printing position matches to zero or
span of the chart.
Pens stand by near the scale center.
Computed correction data are
canceled if the screen is returned to
the operation screen in the procedures
before [Storing].
∗During zero and span adjustments,
operation is not accepted.
: Memory clear
: Adjustment
: Hardware check
: Measured value adjustment
: Shift programming for
measured value
: Printing position adjustment
: Time-axis adjustment for pens
1: 1st pen
2: 2nd pen (2-pen type to 4 pen type)
3: 3rd pen (3-pen type, 4-pen type)
4: 4th pen (4-pen type only)
Store the computed correction data
into memory.
Cursor
Adjustment
Cursor
Printing position
Pen No.
Cursor
(Zero)
2 sec. or
more
Cursor
(Span)
NO
Zero/span adjustment procedure
By returning to the
operation screen
Engineering mode types
Adjustment types
Pen No.
Storing
∗
Other pens than the pen being adjusted
DISP
SHIFT
6
A~Z
°C
°F
ENTRY
°C
°F
ENTRY
ENTRY
ENTRY
°F
°C
ENTRY
DISP
YES
°C
°F
°C
°F
%
/
SHIFT
SET
END
ENTRY
Cursor
Remarks
Ref. 1
Ref. 2
Ref. 3
Ref. 4
Remarks
Reference 1
Reference 2
Reference 3
Reference 4