Installation, Location and external dimensions, Location – Omega RD200 User Manual
Page 13: External dimensions
-9-
2. INSTALLATION
2.1
Location and External Dimensions
Location
Install your recorder at the following place so as not to affect the measuring accuracy and recording
operation unfavorably.
(1) Industrial environment
Select a place being separated from electric field
and magnetic field generating sources and also
free of mechanical vibrations and shocks.
• Overvoltage Category
………………………….
ΙΙ
• Pollution Degree
…………………………………
2
• Altitude
……………………………
2000m or lower
• Working place ……………..………Indoors
(3) Ambient temperature and humidity
Make sure not to expose your recorder to direct
sunlight and not to closely place other materials to
it for preventing rise of its temperature.
• The recommended ambient temperature and
humidity are about 23
°C and about 50%RH.
• Make sure not to expose your recorder to hot
air higher than 70
°C.
• Make sure not place any heat source near to
the terminal board of your recorder.
(2) Atmosphere
Install your recorder in a place where no
inflammable gas exists and no dust, smoke, vapor,
or other dangerous substance exists.
(4) Mounting angle and display view
angle
• Lateral tilting………..
0
° to 10°
• Longitudinal tilting…..Forward tilting: 0
°
Backward tilting: 0 to 30
°
• View angle … -10
° to +30° based on horizontal
External Dimensions
RD200
RD2800
*1: Recorders with the options of MOS relay or ”c” contact alarm output, and communications interface
*2: Recorders with the option of “a” contact alarm output
16
1
□
137
260(274)*1 (285)*2
14
4
18
144
24
63
288
297
288
24
220(236)*1 (247)*2
□
277
*1: Recorders with the options of MOS relay or ”c” contact alarm output, and communications interface
*2: Recorders with the option of “a” contact alarm output
1
2
Display
Approx.
10
°
Approx.
30
°
View angle