beautypg.com

Connections, 4 measuring input terminals – Omega RD200 User Manual

Page 20

background image

-16-

4. CONNECTIONS

4.4 Measuring Input Terminals

1

Measuring input terminals

Make sure to turn off the power supply to prevent
an electric shock.
(1) Measuring input terminals are located on the

down left of the terminal board.

(2) For the connections to the input terminals, use

cables terminated by crimp style terminals with
insulation sleeves.


Input type

Allowable input voltage

Voltage,
Thermocouple input

±10 VDC

(range: ±2V or less)

±60 VDC

(range: ±5V or more)

Resistance

thermometer input

±6 VDC

2

Connections of DC voltage (current) input

Use twisted cables for instrumentation as the
input cables for the purpose of suppressing noise.
For current inputs, mount shunt resistors
(Section 16.2) to the channels to be measured
before connections.

3

Connections of thermocouple inputs

Make sure to use thermocouple wires (or
extension wires) to the input terminals of your
recorder. If a copper wire is used halfway, a
noticeable measuring error occurs. Don’t use a
pair of thermocouple wires in parallel with other
instruments (controller, etc.), otherwise a
malfunction may occur.














4

Connections of resistance thermometer inputs

Use a 3-core cable where each lead wire has an
equal resistance value. Don’t use one resistance
thermometer in parallel with other instruments
(controller, etc.).















A high voltage may be applied to the measuring input terminals due to common mode noise. The
allowable noise value is 30 VAC or less, or 60 VDC or less. Make sure that the noise is lower than the
allowable value. Mount the terminal cover after connections for the purpose of preventing an electric
shock and to protect the input wires. In the case of thermocouple input, the mounting of the terminal cover
can reduce the reference junction compensation error.

Allowable input voltage

mark of measuring input terminals

Terminal

cover

DC voltage input

1 2 3 4

(+)

(-)

Twisted cable for
instrumentation

1 2 3 4

Extension wire

Thermocouple

Red (+)

White (-)

1 2 3 4

Resistance thermometer

3-core cable
(Same diameter,
same length)

Note: Use a 3-core cable where
each lead wire has an equal
diameter and an equal resistance

A

B

B

Caution

Warning

This manual is related to the following products: