18. ADJUSTMENT
18.2 Shift Programming of Measured Value
This programming is for slight-shifting a measured value. After programming, measured values are shifted by the
programmed amount.
1
Shift Programming
(1) This programming is necessary for each channel.
(2) The cursor shifts to the least significant digit. Execute this
programming with or .
2
Programming Flow Chart
Operation screen
[Engineering mode selection]
( + )
3 sec. or more
[Selecting adjustment]
→
[Selecting shift]
→
[Programming channel]
[To measured value display]
[Programming shift]
[To channel programming
screen]
+
Stored
Measured value ….
850.3
Programming ……
850.0
Shift value =
850.0-850.3 = -0.3
Before starting this programming, wait
for at least 30 minutes after turning on
the power supply.
Programmed shifting data is canceled
if the screen is returned to the
operation screen in the procedures
before [Storing].
When a skipped channel is
programmed, operation is not
accepted.
Change the connection to other input
terminals to be adjusted and repeat the
same procedure from the programming
channel display (indicated by a dotted
line).
In zero adjustment or span adjustment
screen, when and are
pressed simultaneously and then
is pressed, the correction data of its
channel are initialized.
: Memory clear
: Adjustment
: Hardware check
: Measured value adjustment
: Shift programming for
measured value
: Printing position adjustment
: Time-axis adjustment for pens
Store the programmed shifting data
into memory.
Cursor Adjustment
Cursor
Shift
Cursor
Channel
Cursor
Channel
Cursor
2 sec. or
more
Cursor
Measured value
Programmed value
Initialization of shift data (0)
Engineering mode types
Adjustment types
Storing
Cursor
Cursor
x 3
Example of shift programming
Before programming
By returning to the
operation screen
Skipped channel does
not accept any [ENTRY]
Other channels adjustment
%
/
%
DISP
SHIFT
6
A - Z
°C
°F
ENTRY
°C
°F
ENTRY
0
CLEAR
1
CHART
ENTRY
%
/
ENTRY
SHIFT
SET
END
ENTRY
0
CLEAR
SHIFT
Remarks 1
Ref. 1
Ref. 2
Ref. 3
Remarks 2
Remarks 3
Remarks 1
Remarks 2
Remarks 3
Remarks 4
ENTRY
Reference 1
Reference 2
Reference 3