beautypg.com

Specifications – Omega RD200 User Manual

Page 132

background image

-128-

22. SPECIFICATIONS

6

Operation Conditions

Operation conditions: Refer to the following table

Items

Reference

operation

Normal operation

Ambient

temperature

21 to 25°C

0 to 31°C

31 to

50°C*

Max. ambient

humidity

80%RH 80%RH

80 to

50%RH*

Min. ambient

humidity

20%RH 20%RH

Power supply

voltage

100VAC

± 1%

90 to 264VAC

Power supply

frequency

50/60Hz

± 0.5%

50/60Hz

± 2%

Left/right

0 to 10°

Forward

tilting

0° 0°

Posture

Backward

tilting

0 to 30°

Warming-up time

30 min. or more

Not necessary

*Decrease from 80% RH at 31

°C, and decrease to 50% at

40

°C

Transportation condition:

Ambient temperature/humidity… -20 to 60

°C, 5 to 90%RH

(without condensation)

Vibration…10 to 60Hz, 4.9 m/S

2

*

Impact…392 m/S

2

*

* When shipping out from a factory
Storage condition:

Ambient temperature/humidity… -20 to 60

°C, 5 to 90%RH

(without condensation)

7

General specifications

Rated power voltage: 100 to 240 VAC, 50/60 Hz
Power consumption: Maximum 60V A
Power failure protection: Programmed parameters stored

into EEPROM memory. Clock circuit sustained for
minimum 8 years by a lithium battery. (8 hours
operation per day)

Insulation resistance:

Between primary and protective conductor terminals

…20M

Ω or more at 500 V DC

Between secondary and protective conductor terminals

…20M

Ω or more at 500 V DC

Between primary and secondary terminals

…20M

Ω or more at 500 V DC

Dielectric strength:

Between primary and protective conductor terminals

…1 minute at 1500 V AC

Between secondary and protective conductor terminals

…1 minute at 500 V AC

Between primary and secondary terminals

…1 minute at 2300 V AC

Primary terminals: Power terminals, Alarm output terminals

(MOS relay, mechanical relay “a” contact)

Secondary terminals: Measuring input terminals, Remote

contacts terminals, Communications interface
terminals , Alarm output terminals (Mechanical relay “c”
contact),

Case assembly material: Door Frame …ABS resin

Front plate…Polycarbonate
Enclosure…Steel

Color: Door … Frame: Black (equivalent to Munsell N3.0)

Front plate: Transparent

Enclosure…Gray (equivalent to Munsell N7.0)

Mounting: Panel mounting
Terminal screws: Power terminals …M4.0

Terminals other than above …M3.5

Weight: RD200 …Approx. 4.0 kg (with fully options)

RD2800 …Approx. 9.0 kg (with fully options)

Dimension: RD200…144Ч144Ч258 (mm)

RD2800…288Ч288Ч220 (mm)

Panel cutout:RD200…138×138 (mm)

RD2800…281×281 (mm)

Clock accuracy: Within

±2 minutes per 30 day

(Under reference operating conditions,
Except errors when power supply is turned
on or off)

Packing material: Layered cardboard is used as cushion

material

8

Safety Regulations

CE: Conforms to EMC and low voltage directives
UL: UL3111-1 (Approval pending)
CSA (C-UL): CSA …C22.2 No.1010 (Approval pending)
IP: Conforms to IEC529 IP54