Printings, 3 list printing – Omega RD200 User Manual
Page 52
-48-
9. PRINTINGS
9.3 List Printing
The details of programming such as "Range/Printing range" and "Scale" can be printed as a list with the plotter
pen. As this includes large amount of data and takes time, it enables to divide the data into two and select printing
of "List 1" or "List 2".
1
List 1 Printing
Operation screen
List 1 printing starts.
Interrupt?
List 1 printing ends.
(Approx. 30 min. for printing)
* With this procedure, the printing operation turns
off. Press and then press again to
turn it on.
(1) RECORD ON status lamp is not lit.
(2) KEY LOCK status lamp lights.
Chart speed does not affect to this operation.
Chart paper forwards at a fixed speed.
(1) Trace printing with a cartridge pen is
interrupted. Other functions are continued
without being interrupted.
(2) Programming any parameters
(Range/Printing range, Scale, Alarm, etc.)
cannot be executed during printing.
Program them after the printing ends.
(1) Press and simultaneously, and then
press .
(2) The selection screen from [List 1] to [List 3] appears.
Move the cursor to “1”.
(3)
When is pressed,
appears for about 5 sec. For execution, go to
procedure
(4).
For no execution, press .
(4) Press to start “List 1” printing. The cartridge
pen moves to the end of the left side and the chart
paper forwards slightly.
(5) For interrupting the printing, press and
simultaneously. Your recorder returns to the normal
operation but the printing operation is turned off. For
turning on the printing operation, press and
then press again.
(6)
When the list printing completes, your recorder
returns to the normal operation.
(1) Year/month/day, Time
(2) Chart speed (Note)
(3) Range/Printing range
(4) Scale
(5) Subtract printing
(6) Periodic data printing
(7) Printing format
(8) Alarm
(Note) When the remote contacts (option), which can
program 3 speeds (No.1 to No.3), is added,
the presently selected speed is printed.
(Note) The above example is for RD200. It is the same for
RD2800.
List 1: Major parameters …. “Range/Printing range”, “Chart speed”, “Printing format”, etc.
List 2: Parameters other than the above …. “Message”, “Time axis synchronization.” and “option-related-items”
List 3: All parameters …. “List 1” + “List 2”
YES*
NO
List 1 printing details
[List No. selection]
Cursor
(1)
(2)
(5)
(6)
List 1 printing cannot be executed in
the following conditions.
Remarks
Operation during printing
Reference
Example of the List 1 printing
DISP
+
or
SHIFT
9
LIST
°C
°F
(3)
(4)
→
ENTRY
ENTRY
→
REC
ON/OFF
ENTRY
ENTRY
DISP
ENTRY
SHIFT
9
LIST
ENTRY
REC
ON/OFF
REC
ON/OFF
ENTRY
REC
ON/OFF
ENTRY
ENTRY