beautypg.com

ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual

Page 362

background image

Maintenance and Troubleshooting 10-41

Problem

Monitor cannot communicate over a serial connection.

Possible

Causes

Serial connection to monitor or field computer may be
loose.

Communication port may not defined correctly in Qstart.

USB Serial Interface cable may be damaged.

Serial port on the computer may not be generating
adequate power for the USB Serial Interface Cable to
facilitate communication with the monitor.

USB drivers are not loaded properly.

3.3-volt fuse on regulator may have blown.

Possible

Solutions

Verify cable connections to monitor and field computer
are secure.

Verify the port is correct through the Control Panel on
your computer.

Inspect USB serial interface cable for damage.

Load USB drivers following the instructions on the USB
cable.

(applies only if the software provides a “Getting Monitor
Attention” message while attempting to communicate
)
Remove the monitor from the hazardous area (at least
20 feet away from the manhole), disconnect the battery
pack, wait at least 1 minute, reconnect the battery pack,
and then attempt to communicate. If the monitor still will
not communicate, replace the 3.3-volt yellow fuse (when
replaceable
) on the regulator.

Contact your regional ADS representative.