Powering up k2 raid – Grass Valley K2 Media Client System Guide v.3.3 User Manual
Page 155

June 9, 2009
K2 Media Client System Guide
155
Powering up K2 RAID
19. In the left pane select
Device Manager
.
20. In the right pane open the
System devices
node and verify that
GVG ISCSI Multi-Path
Device Specific Module
is listed.
21. In Storage Utility, make a new file system
If you get a “…failed to remove the media database…” message, you can safely
proceed.
22. Restart the K2 client and log in to Windows.
23. Open AppCenter and manually remove all clips and bins except the default bin and
the recycle bin.
24. Uninstall and then reinstall both SNFS software and K2 Client software. Use the
sequence and detailed procedure in the K2 Release Notes for the version of K2
Client software currently on the K2 client.
25. As you install K2 Client software, when you arrive at the Specify Target Type
page, select
K2 with local storage
.
26. Restart the K2 client.
27. On a K2 Media Client, from the Windows desktop system tray, open SQL Server
Service Manager and verify that
Auto-start service when OS starts
is selected. (This
is not required on a K2 Media Client.)
The K2 client is now ready for record/play operations.
NOTE: If you ever unbind LUNs, you must do the above procedure again, starting
at step 5.
Powering up K2 RAID
To power up K2 RAID devices, do the following: