Daqcalselectcaltable – Measurement Computing Data Acquisition Systems rev.10.4 User Manual
Page 136

daqCalSelectCalTable
Also See: daqCalGetConstants, daqCalSetConstants,
daqCalSelectInputSignal, daqCalSaveConstants
Format
daqCalSelectCalTable(handle, tableType)
Purpose
daqCalSelectCalTable
selects the calibration table source for the device.
Parameter Summary
Parameter Type
Description
handle DaqHandleT
Handle to the device for which ADC transfer status is to be
retrieved
tableType DaqCalTableTypeT Calibration table type to use
Parameter Values
handle:
obtained from the daqOpen function
tableType:
see table below
Parameter Type Definitions
tableType - DaqCalTableTypeT
Definition
Description
DcttFactory
Selects the factory calibration table. The factory calibration table reflects factory calibration constants for the
selected device. This is the default setting.
DcttUser
Selects the user-calibration table. The user-calibration table reflects calibration constants defined by the user or
the device’s user-calibration application. Refer to the device’s calibration documentation for specific settings.
DcttSelfCal
Self-calibration constants.
Returns
DerrNoError
No error
Function Usage
Selecting the Calibration Table
The daqCalSelectCalTable function should be used to set the current calibration table for the device.
Currently, there are two valid calibration table types which can be selected with the tableType parameter--
either the factory calibration table can be selected with the DcttFactory value, or a user-defined
calibration table can be selected with the DcttUser value. The current calibration table at any time will be
set to the calibration table last selected during the current device session.
Prototypes
C/C++
daqCalSelectCalTable(DaqHandleT handle, DaqCalTableTypeT tableType);
Visual BASIC
VBdaqCalSelectCalTable(ByVal handle&, ByVal tableType)
Program References
None
4.3-10 Daq API Command Reference
908494
Programmer’s
Manual