Fluke Biomedical ProSim 6 User Manual
Page 92

ProSim 6/8
Users Manual
80
Synchronization
Normal Sinus heart rates ................................ 30 to 240 BPM
Maximum rate at 1 ml ................................. 240 BPM achievable with pulses up to 1 ml
Maximum rate at 1.25 ml ............................ 180 BPM
Arrhythmias..................................................... Premature atrial contraction (PAC), Premature ventricular contraction
(PVC), atrial fibrillation, and missed beat.
Leak Test
Target Pressure .................................................. 20 to 400 mmHg
Elapsed time ....................................................... 0:30 to 5:00 minutes:seconds in 30 second steps
Range ................................................................. 0 to 200 mmHg/minute
Internal Leak rate................................................ <2 mmHg/min into 500 ml rigid volume
Pressure Relief Test Range ................................. 100 to 400 mmHg
Oximeter SpO2 Optical Emitter and Detector (optional)
% O2
Range ................................................................. 30 to 100 %
Resolution........................................................... 1 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range..........
±
(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy
With Fluke Biomedical R-curves
91 to 100 %.................................................
±
(3 counts + specified accuracy of the UUT)
81 to 90 %...................................................
±
(5 counts + specified accuracy of the UUT)
71 to 80 %...................................................
±
(7 counts + specified accuracy of the UUT)
Below 71 %................................................. monotonic with unspecified accuracy
Heart Rate
Range ................................................................. 30 to 300 BPM in 1 BPM steps. Oximeter SpO2 Optical Emitter and
Detector is synchronized with ECG rate delayed by 150 ms.
Accuracy ............................................................. ±1 % of setting
Transmission (Ratio of detector current to LED current, expressed in parts per million (ppm))
Range ................................................................. 0 to 300.00 ppm
Resolution........................................................... 0.01 ppm
Accuracy ............................................................. +50 %/-30 % for compatible monitors, unspecified for others. Selected
by finger size and color: Dark, thick finger, medium finger, light, thin
finger, neonatal foot.
Pulse Amplitude
Range ................................................................. 0 to 20.00 %
Resolution........................................................... 0.01 %
Artifact
Respiration
Range ............................................................. 0 to 5 % of transmission
Resolution....................................................... 1 %
Rate ................................................................ all ProSim respiration simulation settings
Ambient Light
Range ............................................................. 0 to 5X transmitted light
Resolution....................................................... 1X
Frequency....................................................... DC, 50 Hz, 60 Hz, and 1 to 10 kHz in 1 kHz steps
Masimo Rainbow Technology ............................. Masimo Rainbow technology with an optional adapter supplied by
Masimo that allows the ProSim two wavelength to test the Rainbow
multiple wavelength system
Compatible Manufacturer Products
With manufacturer R-curve................................. Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke Biomedical R-curve........................... Mindray, GE-Ohmeda, Philips/HP, and BCI