Fluke Biomedical ProSim 6 User Manual
Page 36

ProSim 6/8
Users Manual
24
The rate, amplitude, and width of the QRS wave can be changed.
To change the rate:
1. Push or to highlight the Rate value.
2. Push .
3. Push or to highlight 30, 60, 90, 120, 200, or 250 bpm in the list.
4. Push to set the rate and go back to the ECG: QRS Detection screen.
To change the QRS Wave amplitude:
1. Push or to highlight the Amplitude value.
2. Push .
3. Push or to adjust the amplitude. Each push of a key increases or decreases the
amplitude in the direction of the key pushed. The Product simulates a heart signal
amplitude of 0.05 to 0.25 mV by 0.05 mV intervals and 0.50 to 5.00 mV that is set in
0.25 mV intervals.
4. Push to set the amplitude and go back to the ECG: QRS Detection screen.
To change the width of the QT interval:
1. Push or to highlight the Width value.
5. Push .
6. Push or to adjust the width. Each push of a key increases or decreases the
width 10 ms for each key push when the value is 20 ms or above and 2 ms when the
value is 20 ms or less. The Product simulates a QT width of 8 to 200 ms.
7. When the width is set, push to enter the value and go back to the ECG: QRS
Detection screen.
How to Set Tall T Wave Rejection Test Values (ProSim 8 only)
An ECG monitor must sense and reject a large T wave when it calculates the heart rate.
Use the Tall T Wave Rejection waveform for this test. This wave includes the QRS wave
per EC-13 with a 100 ms width and QT interval of 350 ms. The T wave has a sinusoidal
shape, 180 ms wide, and can be adjusted in amplitude to 0 to 150 % of the overall
waveform amplitude. The rate is set at 80 bpm. To output a tall T wave signal:
1. Push to show the ECG screen in the display.
2. Push or to highlight the Wave Group value.
3. Push .
4. Push or to highlight the Tall T Wave Rejection value in the wave group list.
5. Push to show the ECG: Tall T Wave Rejection in Figure 13 in the display.