Detailed specifications, Normal-sinus-rhythm waveform, Pacemaker waveform – Fluke Biomedical ProSim 6 User Manual
Page 88: Arrhythmia

ProSim 6/8
Users Manual
76
Detailed Specifications
Normal-Sinus-Rhythm Waveform
ECG Reference...................................................... The ECG amplitudes specified are for Lead II (calibration), from the
baseline to the peak of the R wave. All other leads are proportional.
Normal Sinus Rhythm .......................................... 12-lead configuration with independent outputs referenced to right leg
(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.
High-Level Output................................................. 0.5 V/mV
±
5 % of the ECG amplitude setting available on a BNC
connector.
Amplitude .............................................................. 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV
steps). Other leads are proportional to Lead II (reference lead) in
percentage per:
Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 .............................................................. 24
Lead V2 .............................................................. 48
Lead V3 .............................................................. 100
Lead V4 .............................................................. 120
Lead V5 .............................................................. 112
Lead V6 .............................................................. 80
Amplitude Accuracy .............................................
±
(2 % of setting + 0.05 mV)
ECG Rate
ProSim 8 ............................................................. 10 to 360 BPM in 1 BPM steps
ProSim 6 ............................................................. 30 to 360 BPM in 10 BPM steps
Rate Accuracy .......................................................
±
1 % of setting
ECG Waveform Selection..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
ST-Segment Elevation .......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps) Additional steps:
+0.05 mV and -0.05 mV
Power-On Default.................................................. 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Pacemaker Waveform
Pacer-Pulse
Amplitude .......................................................... 0 (off),
±
2,
±
4,
±
6,
±
8,
±
10,
±
12,
±
14,
±
16,
±
18,
±
20,
±
50,
±
100,
±
200,
±
500, and
±
700 mV for lead II (reference lead)
Accuracy
Reference lead II ............................................
±
(5 % setting + 0.2 mV)
All other leads .................................................
±
(10 % setting + 0.4 mV)
Pacer-Pulse Width ................................................ 0.1, 0.2, 0.5, 1.0, and 2.0 ms
±
5 %
Paced Arrhythmias ............................................... Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction
Power-On Default.................................................. Amplitude 10 mV, width 1.0 ms, atrial waveform
Arrhythmia
Baseline NSR......................................................... 80 BPM
PVC Focus ............................................................. Left focus, standard timing (except where specified)
Supraventricular Arrhythmia
ProSim 8 ............................................................. Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed
beat (one time); atrial tachycardia; paroxysmal atrial tachycardia; nodal
rhythm; and supraventricular tachycardia.
ProSim 6 ............................................................. Atrial fibrillation (coarse or fine) and sinus arrhythmia
Premature Arrhythmia