Maintenance
73
•
Do not put battery cells and battery packs near heat or fire.
Do not put in sunlight.
•
Do not disassemble or crush battery cells and battery
packs.
To prevent possible electrical shock, fire, or personal injury:
•
Remove the input signals before you clean the Product.
•
Use only specified replacement parts.
•
Have an approved technician repair the Product.
How to Clean the Product
W Caution
Do not pour fluid onto the Product surface; fluid seepage into
the electrical circuitry may cause the Product to fail.
W Caution
Do not use spray cleaners on the Product; such action may
force the cleaning fluid into the Product and damage electronic
components.
Clean the Product occasionally with a damp cloth and mild detergent. Try to prevent the
entrance of liquids.
Clean the adapter cables with the same precautions. Examine them for damage and
deterioration of the insulation. Examine the connections for integrity. Keep the transducer
adapter clean and dry.
Battery Maintenance
For peak battery performance, charge the Product to maximum charge once a month. If
the Product is not to be used for more than a month, keep it connected to the charger.
Note
To get the specified performance, use the specified battery charger that
comes with this Product.
When the battery gets low, a low battery message shows in the display.
When the battery discharges to 3 % of full charge, a different message shows and the
NIBP function is disabled.
How to Charge the Battery
The battery charge level is shown in the upper-right corner of the display when the
battery pack is installed in the Product. If the Product is plugged in, … shows in the
upper right corner of the display. When the battery charges, the status is updated in
Battery setting under . With the AC/DC power supply removed from the Product,
the battery icon shows the charge level.
The battery can be charged while it is in or out of the Product. The charge rate is slower
when the Product is energized and the battery charger is on. To charge the battery:
1. As shown in Figure 62, connect the AC/DC power supply to the power connector on
the battery pack.
2. Connect the AC/DC power supply to a power source.