You can change the simulated heart rate of the Product from the SpO2 screen. To change
the heart rate:
1. Push or to highlight the Heart Rate value.
2. Push .
3. Push or to adjust the heart rate. Each push of a key moves the heart rate one
beat in the direction of the key pushed. The range is 10 to 300 beats/minute (BPM).
4. Push to go back to the SpO2 screen.
To change SpO2 respiration:
1. From the SpO2 screen, push the or to highlight the Respiration value.
2. Push .
3. Push or to highlight 1%, 2%, 3%, 4%, 5% or off.
4. Push to set the respiration and go back to the SpO2 screen.
You can test SpO2 under different ambient light conditions. To change the ambient light
value:
1. From the SpO2 screen, push the or to highlight the Ambient Light value.
2. Push .
3. Push or to highlight an ambient light value. Below is the list of ambient light
values in the Product.
Sunlight
Indoor – 4 kHz
Indoor – DC
Indoor – 5 kHz
Indoor – 50 Hz
Indoor – 6 kHz
Indoor – 60 Hz
Indoor – 7 kHz
Indoor – 1 kHz
Indoor – 8 kHz
Indoor – 2 kHz
Indoor – 9 kHz
Indoor – 3 kHz
Indoor – 10 kHz
4. Push to go back to the SpO2 screen with the new ambient light value.
The manufacturer must be known before you do a pulse oximeter test, optically through
an artificial finger. For Masimo, you need to know if the sensor is a 2 wavelength or
Rainbow sensor. You can configure the Product for the make of pulse oximeter(s) used
for the test. You change variables for each of the pulse oximeters through the type
variable. To change the oximeter sensor type:
1. From the SpO2 screen, push the or to highlight the Type value.
2. Push .