Fluke Biomedical ProSim 6 User Manual
Page 49

Vital Signs Simulator
How to Simulate Respiration
37
To set the respiration ratio:
1. Push or to highlight the Ratio value.
2. Push .
3. Push or to highlight 1:1, 1:2, 1:3, 1:4, or 1:5.
4. Push to set the respiration ratio and go back to the Respiration screen.
To set the respiration amplitude:
1. Push or to highlight the Amplitude value.
2. Push .
3. Push or to change the amplitude in 0.05
Ω
steps between 0.00 and 5.00
Ω
.
4. Push to set the amplitude and go back to the Respiration screen.
To set the baseline resistance:
1. Push or to highlight the Baseline value.
2. Push .
3. Push or to highlight 500, 1000, 1500, or 2000
Ω
in the list.
4. Push to set the baseline impedance and go back to the Respiration screen.
The baseline impedance is kept in memory and is non-volatile.
Note
Baseline value is kept in memory and is non-volatile.
To set the respiration lead:
1. Push or to highlight the Respiration lead.
2. Push .
3. Push or to highlight LA or LL in the list.
4. Push to set the respiration lead and go back to the Respiration screen. The
respiration lead is kept in memory and is non-volatile.
Note
Respiration lead value is kept in memory and is non-volatile.
How to Set Apnea Simulation
You can simulate an apnea period manually or for a specified time period. To control an
apnea period manually, push the Continuous Apnea softkey from the respiration
screen. The display shows a timer which shows the time since the apnea period started.
The respiration simulation is stopped when the Stop softkey is pushed.
To do a timed apnea period:
1. Push the Timed Apnea softkey from the respiration screen.
2. Push or to highlight 12, 22, or 32 seconds.
3. Push .
The display shows a timer that counts down the apnea period. When the timer counts
down to zero, the apnea period ends and the display shows the respiration screen. Push
the Stop softkey to abort the apnea function.