Invasive blood pressure simulation and tests, How to set the invasive blood pressure variables – Fluke Biomedical ProSim 6 User Manual
Page 41

Vital Signs Simulator
Invasive Blood Pressure Simulation and Tests
29
glh027.bmp
Figure 17. Fetal ECG Intrauterine Pressure Screen
If the contraction value is set to Manual Start, only one contraction cycle is simulated.
The timer shows the time until the contraction ends and the subsequent contraction starts.
Push the Stop softkey to stop contractions and go back to the Fetal ECG: IUP screen.
Invasive Blood Pressure Simulation and Tests
The Product simulates blood pressure for Invasive blood pressure monitors. Each blood
pressure variable can be set through the front-panel controls. The Product also
simulations Swan-Ganz and cardiac catheterization procedures.
How to Set the Invasive Blood Pressure Variables
The Product can simulate two invasive blood pressure transducers at one time. The blood
pressure variables are set separately for each channel. To set these variables, push
to show the IBP screen in Figure 18 in the display.
glh009.bmp
Figure 18. Invasive Blood Pressure Screen
To set the chamber:
1. Push or to highlight the Chamber value on channel 1 or channel 2.
2. Push .
3. Push or to highlight a value in the chamber list. Below is a list of chamber
values in the Product.
Static Right
Ventricle
Arterial Pulmonary
Artery
Radial Artery
PA Wedge
Left Ventricle
Right Atrium (CVP)
Left Atrium