Fluke Biomedical ProSim 6 User Manual
Page 56

ProSim 6/8
Users Manual
44
Note
Heart Rate may not be shown in the NIBP screen if an arrhythmia or
performance waveform is set.
To set the heart rate:
1. Push or to highlight the Heart Rate variable.
2. Push .
3. Push or to increase or decrease the heart rate in 1 bpm steps. The range is 30 to
240 bpm.
Note
When you hold down the direction key, the step size will change to 10 bpm
until the key is released.
4. Push to set the heart rate and go back to the NIBP screen.
Note
Arrhythmia may show in the display instead of Heart Rate when an
arrhythmia or performance waveform is set under the ECG function. The
Mono VTach waveform is the one exception. You can only change the heart
beat to normal sinus rhythm in the NIBP screen.
To set the heart beat waveform to normal sinus rhythm:
1. Push or to highlight the Arrhythmia variable.
2. Push .
3. Push or to increase or decrease the heart rate.
4. Push to set the heart rate and go back to the NIBP screen.
To set the pulse volume:
1. Push or to highlight the Pulse Volume variable.
2. Push .
3. Push or to increase or decrease the pulse volume in 0.05 mL steps. The pulse
volume range is 0.10 to 1.25 mL.
4. Push to set the pulse volume and go back to the NIBP screen.
You can set the wave variable to Adult or Neonatal simulation. When set, the pulse
volume is set to the default for that simulation: 1.0 ml for Adult, 0.5 ml for Neonatal.
Afterwards, the pulse volume can be changed with its own control (see above). To set the
wave variable:
1. Push or to highlight the Wave variable.
2. Push .
3. Push or to highlight Adult or Neonatal.
4. Push to set the wave and go back to the NIBP screen.
When the NIBP monitor starts the blood pressure measurement cycle, the Product keeps
the measured parameters of the measurement cycle. If the NIBP screen is shown in the
display when the measurement cycle starts, a graph of pressure versus time shows in the
display. See Figure 35.