How to turn the product on, How to connect a pc to the product – Fluke Biomedical ProSim 6 User Manual
Page 19

Vital Signs Simulator
How to Turn the Product On
7
How to Turn the Product On
Push on the front panel to turn the Product on. The screen shown in Figure 3 shows
the power-up screen.
glh002.bmp
Figure 3. Power-Up Screen
When the self test is complete and no errors are sensed, the home screen in Figure 4
shows in the display.
glh001.bmp
Figure 4. Home Screen
From the home screen, a number of pre-defined patient simulations can be used to set all
the simulation functions at one time. See Pre-Defined Simulations to learn more on how
to use simulations. Waveform and ECG rate parameters are set through the ECG
function. See the ECG Function section to learn more on how to set these variables. The
respiration rate (Resp Rate), temperature parameter, cardiac output, and fetal simulation
are set through the special functions menu. The two Invasive Blood Pressure (IBP)
channels are set through the IBP function. The Non-Invasive Blood Pressure variables are
set through the NIBP function. See the Blood Pressure Simulations and Tests section to
learn more on how to set IBP and NIBP variables. The SpO2 variable is set through the
SpO2 function. See the Oximeter SpO2 Optical Emitter and Detector section to learn
more on how to set this variable.
How to Connect a PC to the Product
You can use a PC to store presets, auto sequences, and test results from the Product. You
must use the ProSim Mini Plug-In on the PC to communicate with the Product.
Note
You must have Ansur Test Executive version 2.9.6 on your PC to
communicate with the Product.