Fluke Biomedical ProSim 6 User Manual
Page 22

ProSim 6/8
Users Manual
10
Table 7. Pre-Defined Patient Simulations (cont.)
Simulation Name
Parameter
Pre-Set Value
Wave Form
NSR (Adult)
ECG Rate
30 bpm
Respiration Rate
15 brpm
Temperature 35.0
°
C
IBP Channel 1
100/65 mmHg (Art)
IBP Channel 2
25/10 mmHg (PA)
NIBP
100/65 (77) mmHg
Bradycardic
SpO2 85%
Wave Form
Atrial Fibrillation (Coarse)
ECG Rate
No bpm
Respiration Rate
70 brpm
Temperature 34.0
°
C
IBP Channel 1
35/15 mmHg (Art)
IBP Channel 2
35/0 mmHg (LV)
NIBP
35/15 (22) mmHg
Ventricular Fibrillition
SpO2 80
%
Wave Form
Asystole
ECG Rate
0 bpm
Respiration Rate
0 brpm
Temperature 33.0
°
C
IBP Channel 1
0 mmHg
IBP Channel 2
0 mmHg
NIBP 0/0
mmHg
Asystole
SpO2 50
%
After you set one of the pre-defined simulations, you can change the parameters of one or
all of the vital signs. Changes are not kept in the Product. The pre-defined simulation
parameters can not be changed through the front panel.
There are seven custom pre-defined simulations in the Product. To set a custom
simulation:
1. From the Home screen, push the More softkey.
2. Push the Custom softkey.
Four custom simulation names show above the first four softkeys. The names are
Custom1, Custom2, Custom3, and Custom4. To show the next three names,
push the More softkey.
3. Push the softkey for a custom pre-defined simulation.