How to make an operator id – Fluke Biomedical ProSim 6 User Manual
Page 75

Vital Signs Simulator
How to Save and View Test Results (ProSim 8 only)
63
Operator: John Doe
Test Record ID: ABC123 1/20/2011
Test Results 001
Test Results 002
Test Results 003
•
•
•
Test Record ID: XYZ987 1/31/2011
Test Results 001
Test Results 002
Test Results 003
•
•
•
Operator: Jane Doe
Test Record ID: ABC123 1/15/2011
Test Results 001
Test Results 002
Test Results 003
•
•
•
Test Record ID: DEF456 1/15/2011
Test Results 001
Test Results 002
Test Results 003
•
•
•
glh044.eps
Figure 52. Memory Organization of Test Results
How to Make an Operator ID
Test results are kept in memory by an operator ID. To make a new operator ID:
1. Push .
2. Push or to highlight Test Record ID in the list.
3. Push to open the Test ID screen shown in Figure 53.
glh036.bmp
Figure 53. Operator ID Screen
4. Push the Change Operator softkey to open the change operator screen shown in
Figure 54.