Fluke Biomedical ProSim 6 User Manual
Page 32

ProSim 6/8
Users Manual
20
2. Push .
3. Push or to highlight an amplitude value in the list. Below is a list of the
pacemaker amplitude values.
0 mV
16 mV
2 mV
18 mV
4 mV
20 mV
6 mV
50 mV
8 mV
100 mV
10 mV
200 mV
12 mV
500 mV
14 mV
700 mV
4. Push to set the amplitude and go back to the TV Paced Settings screen.
To change the width value:
1. Push or to highlight the Width value.
2. Push .
3. Push or to highlight 0.1, 0.2, 0.5, 1.0, 1.5, or 2.0 mV in the list.
4. Push to set the width and go back to the TV Paced Settings screen.
To change the polarity:
1. Push or to highlight the Polarity value.
2. Push .
3. Push or to highlight + or - in the list.
4. Push to set the polarity and go back to the TV Paced Settings screen.
Push the Back softkey to go back to the ECG: TV Paced screen.
Advance Cardiac Life Support (ACLS) Waveforms (ProSim 8 only)
The Product can simulate Advanced Cardiac Life Support (ACLS) waveforms.
To simulate an ACLS waveform:
1. From the ECG screen, push or to highlight the Wave Group value.
2. Push .
3. Push or to highlight ACLS in the wave group list.
4. Push .
5. Push or to highlight the Waveform value.
6. Push .
7. Push or to highlight a waveform in the list.
8. Push to set the waveform and go back to the ECG: ACLS screen.
The amplitude of all ACLS waveforms can be adjusted. See the How to Set the ECG
Parameters section above. ECG signal artifacts can also be added to each arrhythmia
signal. See the How to Simulate an ECG Artifact section above.