Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM
R-Wave up slope ................................................ 0.875 amplitude, 0.4375 x width
R-Wave down slope............................................ Full amplitude, 0.5 x width
S-Wave up slope ................................................ 0.125 amplitude, 0.0625 x width
Tall T-Wave Rejection (ProSim 8 only)
Waveform
QT Interval ...................................................... 350 ms
T-Wave width.................................................. 180 ms
T-Wave Shape................................................ ½ sinewave
Amplitude............................................................ 0 to 150 % reference lead amplitude in 10 % steps
Rate .................................................................... 80 BPM
Rate Accuracy .......................................................
±
1 % of setting
Amplitude Accuracy .............................................
±
(2 % of setting + 0.05 mV)
ECG Artifact
Type........................................................................ 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size......................................................................... 25, 50, 100 % of the normal sinus R-Wave for each lead
Lead Select ............................................................ All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal / Maternal-ECG (ProSim 8 only)
Fetal Heart Rate (Fixed)........................................ 60 to 240 BPM in 1 BPM steps
Fetal Heart Rate (IUP) ........................................... 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ....................... Early deceleration, late deceleration, and acceleration
Wave Duration....................................................... 90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and
returning to 0
IUP Period.............................................................. 2, 3, or 5 minutes; and manual
Default Settings..................................................... FHR 140 BPM, early deceleration wave, manual
Invasive Blood Pressure
Channels................................................................ 2, each independently settable with identical parameters and are
individually electrically isolated from all other signals
Input/output Impedance ....................................... 300
Ω
±
10 %
Exciter Input Range .............................................. 2.0 to 16.0 V peak
Exciter-Input Frequency Range........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5 (default) or 40
μ
V/V/mmHg
Pressure Accuracy ...............................................
±
(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation
only
Static Pressure...................................................... -10 to +300 mmHg in 1 mmHg steps
Pressure Units....................................................... mmHg or Kpa
Dynamic Waveforms
Types (default pressures) ................................... Arterial (120/80)
Radial artery (120/80)
Left ventricle (120/00)
Right ventricle (25/00)
Pulmonary artery (25/10)
Pulmonary-artery wedge (10/2)
Right atrium (central venous or CVP) (15/10)
Pressure Variability............................................. Systolic and diastolic pressures are independently variable in 1 mmHg
steps.
Swan-Ganz Sequence........................................... Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary
artery wedge (PAW)
Cardiac Catheterization (ProSim 8 only)
Chambers ........................................................... Aortic, Pulmonary valve, and Mitral valve
Respiration Artifact
Arterial, radial artery, and left ventricle ............... 5 to 10 % multiplication
Other ................................................................... 5 or 10 mmHg
BP Output .............................................................. Circular DIN 5-Pin
Power-On Default.................................................. 0 mmHg