Fluke Biomedical ProSim 6 User Manual
Page 62

ProSim 6/8
Users Manual
50
ProSim 8
Mandrel
Blood Pressure Cuff
Blood Pressure
Gauge
Wraps around mandrel.
glh051.eps
Figure 42. Pressure Source Connection
2. From the NIBP screen, push the Tests softkey.
3. Push or to highlight Pressure Source.
4. Push .
5. Push or to increase or decrease the target pressure between 20 and 400 mmHg
in 1 mmHg steps.
Note
When you hold down the direction key, the step size will change to
10 mmHg until the key is released.
6. Push the Start softkey to begin the test.
The Product pressurizes the pneumatic system to the target pressure. While the Product
pumps air into the pneumatic system, the pressure measurement and a graph of the
pressure is shown in the display. See Figure 43.
glh010.bmp
Figure 43. Pressure Source Pumping Screen