How to set battery settings, How to set the display language – Fluke Biomedical ProSim 6 User Manual
Page 83

Vital Signs Simulator
Setup Features
71
4. Push to set the sensitivity and go back to the Pressure Settings screen.
To set the Pressure units:
1. Push or to highlight the Units value.
2. Push .
3. Push or to highlight mmHg or kPa.
4. Push to set the units and go back to the IBP Settings screen.
How to Set Battery Settings
To help save battery life, the Product can be set to power down (Auto power off) when no
buttons are sensed as pushed for a set period of time. You can also set whether or not the
battery charges in the Product. From the Setup screen, push or to highlight
Battery Settings and push . To go back to the Setup screen, push the Back
softkey.
To set auto power off:
1. Push or to highlight the Auto Power Off value.
2. Push .
3. Push or to highlight an auto power off value.
When enabled, auto power of can be set to 10, 30, or 60 minutes.
4. Push to set auto power and go back to the Battery Settings screen.
To set the battery charger:
1. Push or to highlight the Charge Battery value.
2. Push .
3. Push or to highlight Yes or No.
4. Push to set the charger value and go back to the Battery Settings screen.
Note
A 1-minute warning message will show in the display before the Product
turns off.
How to Set the Display Language
The Product can be set to show text and messages in the display in many different
languages. To set the language, push or to highlight Instrument Information in
the Setup screen and push . To go back to the Setup screen, push the Back
softkey.
The language already set in the Product, shows in the display. To change the language:
1. Push .
2. Push or to highlight a language in the list. Below is a list of the display
languages.
English Spanish
French Japanese
Italian Chinese
German
3. Push .