Fluke Biomedical ProSim 6 User Manual
Page 42

ProSim 6/8
Users Manual
30
4. Push to set the chamber and go back to the IBP screen.
You must first zero the patient monitor when you connect to a new IBP transducer or
patient simulator. To set a channel to 0 mmHg and Static chamber, push the Zero
Channel 1 or Zero Channel 2 softkey. After you zero the product, do the zero
function on the patient monitor.
To set the pressure:
1. Push or to highlight the Pressure value on channel 1 or channel 2.
2. Push .
Systolic and diastolic pressures are set separately.
Note
If the chamber is set to static, only one pressure is set.
3. Push or to highlight the systolic or diastolic pressure value.
4. Push or to increase or decrease the pressure value in 1 mmHg steps. The range
is -10 to +300 mmHg.
Note
When you hold down the direction key, the step size will change to
10 mmHg until the key is released.
5. Push to set the pressure and go back to the IBP screen.
When the Chamber parameter is set to a value other than Static, the artifact parameter is
added to the display as shown in Figure 19.
glh051.eps
Figure 19. IBP with Chambers Screen
To set the artifact:
Note
Artifact value is not available when the chamber value is set to static.
1. Push or to highlight the Artifact value on channel 1 or channel 2.
2. Push .
3. Push or to highlight 5% or 10% if the chamber parameter is set to Arterial,
Radial Artery, or Left Ventricle. Highlight 5 mmHg or 10 mmHg for all other
chamber values. Highlight Off to turn off the artifact function