How to set the cardiac output waveform, How to set the baseline temperature, How to set injectate temperature – Fluke Biomedical ProSim 6 User Manual
Page 53: How to start a cardiac output simulation, Non-invasive blood pressure simulation and tests

Vital Signs Simulator
Non-Invasive Blood Pressure Simulation and Tests
41
How to Set the Cardiac Output Waveform
To set the cardiac output waveform:
1. In the Cardiac Output screen, use or to highlight the wave parameter.
2. Push .
3. Push or to highlight a waveform name in the list of waveforms.
Table 13 shows the cardiac output waveforms for the Product.
Table 13. Cardiac Output Waveforms
Waveform Description
2.5 L/min
Normal waveform with accuracy of 2.5 L/min
5.0 L/min
Normal waveform with accuracy of 5.0 L/min
10.0 L/min
Normal waveform with accuracy of 10.0 L/min
Interrupted Injectable
Interrupted injection waveform
LR Shunt
Left to right ventricular shunt
Calibrated Pulse
Calibrated square wave pulse
4. Push to set the wave and go back to the Cardiac Output screen.
How to Set the Baseline Temperature
To set the baseline temperature:
1. In the cardiac output screen, if not already highlighted, use or to highlight the
baseline temperature value.
2. Push .
3. Push or to highlight 36
°
C, 37
°
C, or 38
°
C.
4. Push to set the baseline temperature and go back to the Cardiac Output
screen.
How to Set Injectate Temperature
To set the injectate temperature:
1. In the cardiac output screen, use or to highlight the injectate temperature value.
2. Push .
3. Push or to highlight 0
°
C or 24
°
C.
4. Push to set the temperature and go back to the Cardiac Output screen.
As you change the injectate temperature, the calibration coefficient necessary for the
monitor is shown in the display.
How to Start a Cardiac Output Simulation
After you set the parameters for a cardiac output simulation, push the Start softkey. The
simulation completes automatically. To stop the simulation, push the Stop softkey.
Non-Invasive Blood Pressure Simulation and Tests
The Product simulates blood pressure for Non-Invasive blood pressure monitors. Each
blood pressure variable can be set through the front-panel controls. The Product also does
leak, pressure source, and pressure relief tests. The manometer function sets the Product
to measure static pressure and shows the pressure in the display.