Autosequences
61
Table 14. Autosequences (cont.)
Autosequence
Sequence Steps
Run Time
Waveform: Triangle; Rate: 2.0 Hz; Amplitude: 1.0 mV
00:05
Waveform: Pulse; Rate: 60 bpm; Amplitude: 1.0 mV
00:05
NIBP Sequence (cont.)
REPEAT
Total time 01:10
IBP 1: 120/80 mmHg (Art), 2: 120/0 mmHg (LV)
01:00
IBP 1: 25/0 mmHg (RV), 2: 120/80 mmHg (Art)
01:00
IBP 1: 25/10 mmHg (PA), 2: 10/2 mmHg (PAW)
01:00
IBP Sequence
REPEAT Total
Time
03:00
Temperature 30.0
°
C 00:20
Temperature 34.0
°
C 00:20
Temperature 37.0
°
C 00:20
Temperature 42.0
°
C 00:20
Temperature Sequence
STOP Total
Time
01:20
How to View the Steps of an Autosequence
To view an autosequence:
1. Push .
2. Push , , , or to highlight Autosquences.
3. Push .
4. Push or to highlight an autosequence in the list. See Table 14.
5. Push .
The screen shown in Figure 50 is the Temperature Sequence in the autosequence list.
glh032.eps
Figure 50. Autosequence Start Screen
The details of the autosequence shows the sequence does each of the four steps and stops.
Although this example does not repeat, some autosequences do. The screen also shows
the four-step sequence will complete in 1 minute and 20 seconds.
Each sequence step is shown in the display. When there are more steps than can be
shown in one display screen, push or to go to the next or previous screen.