How to simulate respiration – Fluke Biomedical ProSim 6 User Manual
Page 48

ProSim 6/8
Users Manual
36
To set the simulated temperature:
1. Push .
2. Push , , , or to highlight the Temperature selection then push to
show the set temperature in the display.
3. Push .
4. Push or to change the temperature in 0.5
°
C steps between 30.0 and 42.0
°
C.
Note
When you hold down the direction key, the step size will change to 1.0
°
C
until the key is released.
5. Push to set the temperature and go back to the Temperature screen.
How to Simulate Respiration
Respiration variables are set through the special functions. To set respiration:
1. Push .
2. Push , , , or to highlight Respiration.
3. Push . The Respiration screen in Figure 27, shows in the display.
glh013.bmp
Figure 27. Respiration Screen
You can choose between a normal or ventilated respiration waveform and change the
respiration rate. The respiration signal can be set to the left arm (LA) or left leg (LL)
ECG lead. The baseline impedance between the leads and the amplitude of impedance
variation (respiration amplitude) are set though the front panel as well. To set respiration
between normal and ventilated:
1. Push or to highlight the Wave value.
2. Push .
3. Push or to highlight Normal or Ventilated.
4. Push to set the wave and go back to the Respiration screen.
To set the respiration rate:
1. Push or to highlight the Rate value.
2. Push .
3. Push or to increase or decrease the rate of respiration in 1 brpm steps between
10 and 150 brpm.
4. Push to set the respiration rate and go back to the Respiration screen.