Fluke Biomedical ProSim 6 User Manual
Page 59

Vital Signs Simulator
Non-Invasive Blood Pressure Simulation and Tests
47
Note
If the NIBP device has an internal system leak test or one that vents the cuff
inflation pneumatic circuit to the atmosphere when idle, do not use the Leak
Test. Rather, do a Manometer check to test for internal system leaks. Refer
to the NIBP monitor operators manual for the recommended test protocol.
To perform a leak test:
1. Push to show the NIBP screen.
2. Push the Tests softkey.
3. Push or to highlight Leak Test.
4. Push to show the NIBP: Leak Test screen in Figure 38 in the display.
glh024.bmp
Figure 38. Leak Test Screen
The default target pressure is 300 mmHg. The target pressure can be set between 20
and 400 mmHg. To change the target pressure, push or to increase or decrease
the pressure value in 1 mmHg steps.
Note
When you hold down the direction key, the step size will change to
10 mmHg until the key is released.
5. Push the Start softkey to start the leak test.
The Product pressurizes the pneumatic system to the target pressure. While the
Product pumps air into the pneumatic system, the pressure and a graph of the
pressure is shown in the display. See Figure 39.
glh006.bmp
Figure 39. Leak Test Pumping Screen