Respiration, Temperature, Cardiac output – Fluke Biomedical ProSim 6 User Manual
Page 91: Non-invasive blood pressure

Vital Signs Simulator
Detailed Specifications
79
Respiration
Rate ........................................................................ 0 (OFF), 10 to 150 BrPM in 1 BrPM steps
Waves..................................................................... Normal or ventilated
Ratio (inspiration:expiration)
Normal ................................................................ 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated............................................................ 1:1
Impedance Variations (
Δ
Ω
) ................................. 0.00 to 1.00
Ω
in 0.05
Ω
steps and 1.00 to 5.00
Ω
in 0.25
Ω
steps
Delta Accuracy ......................................................
±
(5 % of setting + 0.1
Ω
)
Baseline ................................................................. 500, 1000 (default), 1500, 2000
Ω
, Leads I, II, III
Baseline Accuracy ................................................
±
5 %
Respiration Lead................................................... LA or LL (default)
Apnea Selection .................................................... 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON =
respiration OFF)
Power-On Default.................................................. 20 BrPM, delta 1.0
Ω
Temperature
Temperature .......................................................... 30.0
°
C to 42.0
°
C in 0.5
°
C steps
Accuracy................................................................
±
0.4
°
C
Compatibility ......................................................... Yellow Springs, Inc. (YSI) Series 400 and 700
Output .................................................................... Circular DIN 4-pin
Cardiac Output
Catheter Type ........................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient.......................................... 0.542 (0
°
C injectate), 0.595 (24
°
C injectate)
Blood Temperature ............................................... 36
°
C (98.6
°
F) to 38
°
C (100.4
°
F)
±
0.2
°
C in 1
°
C steps
Injectate Volume ................................................... 10 cc
Injectate Temperature........................................... 0
°
C or 24
°
C
Cardiac Output ...................................................... 2.5, 5, 10 liters per minute
±
7.5 %
Faulty-Injectate Curve .......................................... Waveform for simulation available
Left-to-Right-Shunt Curve.................................... Waveform for simulation available
Calibrated Pulse.................................................... 1.5
°
for 1 second
Connector .............................................................. Circular DIN 7 pin
Power-On Default ................................................. 5 liters per minute, 0
°
C injectate, 37
°
C blood temperature
Non-Invasive Blood Pressure
Pressure Units....................................................... mmHg or kPa
Manometer (Pressure Meter)
Range ................................................................. 10 to 400 mmHg
Resolution........................................................... 0.1 mmHg
Accuracy
ProSim 8 .........................................................
±
(0.5 % reading +0.5 mmHg)
ProSim 6 .........................................................
±
(1 % reading +1 mmHg)
Pressure Source
Target Pressure Range ...................................... 20 to 400 mmHg
Resolution........................................................... 1 mmHg
NIBP Simulations
Pulse................................................................... 2 mmHg max into 500 ml NIBP system
Volume of air moved........................................... 1.25 ml max
Simulations ......................................................... Systolic/diastolic (MAP)
Adult................................................................ 60/30 (40), 80/50 (60), 100/65 (77); 120/80 (93); 150/100 (117); and
200/150 (167) and 255/195 (215)
Neonatal ......................................................... 35/15 (22); 60/30 (40); 80/50 (60);100/65 (77);120/80 (93) and 150/100
(117)
Pressure variability ......................................... Systolic and diastolic pressures are variable by 1 mmHg
Repeatability ....................................................... Within
±
2 mmHg (at maximum pulse size independent of device under
test)