Vital Signs Simulator
Autosequences
59
Table 14. Autosequences (cont.)
Autosequence
Sequence Steps
Run Time
ECG 80 bpm, IBP1 120/80 (Art), IBP2 120/0 (LV), SpO2 
98 %, NIBP 120/80, Respiration 20 brpm, and 
Temperature 37 
°
C
01:00
ECG 120 bpm, IBP1 250/150 (Art), IBP2 200/0 (LV), 
SpO2 85 %, NIBP 200/150, Respiration 60 brpm, and 
Temperature 40 
°
C
01:00
ECG 30 bpm, IBP1 60/30 (Art), IBP2 60/0 (LV), SpO2 
50 %, NIBP 60/30, Respiration 10 brpm, and 
Temperature 34 
°
C
01:00
Medical Training
REPEAT Total
Time
03:00
SpO2 100 %, Pulse Amplitude 20 %, Transmissivity: 
normal adult, heart rate 60BPM 
00:45
SpO2 95 %, Pulse Amplitude 15 %, Transmissivity: 
normal adult, heart rate 60BPM 
00:45
SpO2 90 %, Pulse Amplitude 10 %, Transmissivity: 
dark/thick adult, heart rate 120BPM 
00:45
SpO2 85 %, Pulse Amplitude 7 %, Transmissivity: 
dark/thick adult, heart rate 30BPM 
00:45
SpO2 84 %, Pulse Amplitude 5 %, Transmissivity: 
normal adult, heart rate 60BPM 
00:45
SpO2 83 %, Pulse Amplitude 4 %, Transmissivity: 
normal adult, heart rate 60BPM 
00:45
SpO2 82 %, Pulse Amplitude 3 %, Transmissivity: 
dark/thick adult, heart rate 120BPM 
00:45
SpO2 81 %, Pulse Amplitude 2 %, Transmissivity: 
dark/thick adult, heart rate 30BPM 
00:45
SpO2 80 %, Pulse Amplitude 1 %, Transmissivity: 
normal adult, heart rate 60BPM 
00:45
SpO2 79 %, Pulse Amplitude 0.5 %, Transmissivity: 
normal adult, heart rate 60BPM 
00.45
Oximeter Testing
STOP
Total time 07:30
NSR (Adult) 80 BPM
00:45
Premature PVC1 Left Vent
00:30
Ventricular Mono VTach
00:30
Ventricular Ventricular Fibrillation
00:30
Ventricular Asystole
00:25
Cardiac Failure
STOP Total
Time
02:30