Fluke Biomedical ProSim 6 User Manual
Page 55

Vital Signs Simulator
Non-Invasive Blood Pressure Simulation and Tests
43
Figure 33 shows the neonatal blood pressure cuff mandrel.
Large
Medi
u
m
Small
fcv012.eps
Figure 33. Neonatal Blood Pressure Cuff Mandrel
To set the blood pressure simulation, push to show the NIBP screen in Figure 34 in
the display.
glh003.bmp
Figure 34. Non-Invasive Blood Pressure Screen
Pressure, heart rate, pulse volume, brand, and wave are set through the front-panel
controls to simulate different patient conditions. Arrhythmia waveforms are also
simulated in the NIBP simulation (set through ECG). To set the Blood Pressure:
1. Push or to highlight the Pressure variable.
2. Push .
Systolic and diastolic pressures are set separately.
3. Push or to highlight the systolic or diastolic pressure value.
4. Push or to increase or decrease the pressure value in 1 mmHg steps.
Note
When you hold down the direction key, the step size will change to
10 mmHg until the key is released.
5. Push to set the pressure and go back to the NIBP screen.