How to view test results – Fluke Biomedical ProSim 6 User Manual
Page 78

ProSim 6/8
Users Manual
66
3. Push or to set the parameter to the value observed on the UUT.
4. Push to set the observed value and show the save screen shown below in the
display.
glh049.bmp
Figure 56. Observed Results Screen
Note
If the observed value is incorrect at this point, you can push the Enter
Observed softkey and change the observed parameter.
5. Test results are saved with a Pass or Fail prefix in their label. Push the Save as
Pass or the Save as Fail softkey. The saving screen in Figure 57 shows in the
display.
glh040.bmp
Figure 57. Saving Screen
The three-digit number is the test result ID. Within the test ID, each saved test results
is assigned a number in sequence. When the save operation completes, the display
goes back to the screen for the function you just saved. In the case of this example,
the display goes back to the temperature screen.
How to View Test Results
To view the test results in memory:
1. Push .
2. Push , , , and to highlight View Memory.
3. Push .
Note
If no test results data is saved to memory, an error message shows in the
display when you push .