Fluke Biomedical ProSim 6 User Manual
Page 21

Vital Signs Simulator
Pre-Defined Simulations
9
Table 7. Pre-Defined Patient Simulations
Simulation Name
Parameter
Pre-Set Value
Wave Form
NSR (Adult)
ECG Rate
60 bpm
Respiration Rate
20 brpm
Temperature 37.0
°
C
IBP Channel 1
120/80 mmHg (Art)
IBP Channel 2
28/15 mmHg (PA)
NIBP
120/80 (93) mmHg
Normal
SpO2 97
%
Wave Form
NSR (Adult)
ECG Rate
130 bpm
Respiration Rate
40 brpm
Temperature 37.0
°
C
IBP Channel 1
200/150 mmHg (Art)
IBP Channel 2
45/25 mmHg (PA)
NIBP 200/150
(166)
mmHg
Hypertensive
SpO2 94
%
Wave Form
NSR (Adult)
ECG Rate
40 bpm
Respiration Rate
15 brpm
Temperature 37.0
°
C
IBP Channel 1
60/30 mmHg (Art)
IBP Channel 2
15/10 mmHg (PA)
NIBP
60/30 (40) mmHg
Hypotensive
SpO2 95
%
Wave Form
NSR (Adult)
ECG Rate
180 bpm
Respiration Rate
50 brpm
Temperature 40.0
°
C
IBP Channel 1
80/50 mmHg (Art)
IBP Channel 2
25/10 mmHg (PA)
NIBP
80/50 (60) mmHg
Tachycardia
SpO2 88
%