How to set the spo2 parameters – Fluke Biomedical ProSim 6 User Manual
Page 67

Vital Signs Simulator
Oximeter SpO2 Optical Emitter and Detector
55
ProSim 8
SpO
2
Finger Module
glh045.eps
Figure 49. Magnetic Holder for SpO2 Artificial Finger
How to Set the SpO2 Parameters
You can raise or lower the degree of oxygen saturation. To change the SpO2 value:
1. Push or to highlight the SpO2 value.
2. Push .
3. Push or to increase or decrease the SpO2 value in 1 % steps between 30 and
100 %.
4. Push to set that SpO2 value and go back to the SpO2 screen.
To change the pulse amplitude value:
1. Push or to highlight the Pulse Amplitude value.
2. Push .
3. Push or to change the pulse amplitude value between 0 and 20 %. Each push of
a key changes the value 0.01 % between 0 and 1 % in the direction of the key pushed.
The value changes 0.1 % between 1 and 10 %. The value changes 1 % between 10
and 20 %.
4. Push to set the pulse amplitude value and go back to the SpO2 screen.
To change the transmission value:
1. From the SpO2 screen, push or to highlight the Transmission value.
2. Push .
3. Push or to highlight Dark, Thick Finger, Medium Finger, Light, Thin
Finger, and Neonatal Foot.
4. Push to set that transmission value and go back to the SpO2 screen.