How to do a pressure leak test – Fluke Biomedical ProSim 6 User Manual
Page 58

ProSim 6/8
Users Manual
46
glh053.bmp
Figure 36. NIBP Summary Screen
4. Compare the NIBP monitor values with the target values shown in the Product
display.
How to Do a Pressure Leak Test
The leak test measures leaks in a non-invasive blood pressure monitor, the hoses
connected to the monitor, and the pressure cuff.
Note
Before you do a pressure leak test on a monitor, do the pressure leak test
without the monitor to identify the leak rate of the Product. Use this leak
rate to offset the rate of the full system with the monitor connected.
Note
Put the NIBP monitor in “calibrate” or “service” mode to close the vent
valve, so the Product can inflate the pneumatic system. Refer to the service
manual for the NIBP monitor.
Connect the Product to the monitor and cuff as shown in Figure 37.
ProSim 8
Patient
Monitor
Mandrel
Dual hose system: connect Cuff Adapter to
hose marked “Sense”. If both hoses are
unmarked, connect Cuff Adapter to either
hose.
Blood Pressure Cuff
Wraps around mandrel.
Must be connected closer
to the cuff than monitor.
glh008.bmp
Figure 37. Pressure Leak Test Connections