How to make a test id – Fluke Biomedical ProSim 6 User Manual
Page 76

ProSim 6/8
Users Manual
64
glh037.bmp
Figure 54. Change Operator ID Screen
5. Push , , , and to highlight a character for the operator ID.
6. Push to add the highlighted character to the operator ID.
7. Do steps 5 and 6 for each character in the operator ID. You can have the maximum
number of characters that fit in the test ID field in the display. If you use more of the
narrow characters like I and 1, you can have more characters in the operator ID than
when wider characters are used.
8. Push the Done softkey when you have all the characters in the operator ID.
Note
If this is a new operator ID, the ID won’t exist until you save the first test
result.
Do these same steps when you change the operator ID to an ID that is already in memory.
After you type the ID in as it already is in memory, all saved results data are added to that
operator ID.
How to Make a Test ID
As shown in Figure 52, test results data is related to a Test Record ID. This test record ID
could identify the device under test. Some examples could be its model number, its serial
number, or its asset number. After you input the new Test ID, all results data saved after
that point is kept with that test ID until you change the test ID.
Note
When you type in a new test ID, the test ID that was open before is closed
and no more test results data can be added to that test ID. You can use the
same test ID name as one that is already in memory, but a date and time
stamp will always be appended to the name and its own test ID.
To make a test ID:
Note
These instructions use the setup function to make a test ID. You can also
make a test ID when a Save softkey label shows in the display. The New
Test ID softkey always shows in the display when you push the Save
softkey.
1. Push .
2. Push or to highlight Test Record ID in the list.
3. Push to open the Test ID screen.
4. Push the New Test ID softkey.