To change the ECG waveform:
1. Push or to move the highlight to the Wave Group value.
2. Push .
3. Push or to move the highlight to a waveform group name. Table 8 is a list of
the wave groups available in the Product.
Table 8. ECG Wave Groups
Wave group
NSR (Adult)
Adult Normal Sinus Rhythm
NSR (Pediatric)
Child Normal Sinus Rhythm
Performance
R-Wave Detection (ProSim 8 only)
QRS Detection (ProSim 8 only)
Tall T Wave Rejection (ProSim 8 only)
Special waveforms for monitor tests. See the ECG
Tests section.
Supraventricular
Premature
Ventricular
Conduction
TV Paced
ACLS (ProSim 8 only)
Arrhythmia waveforms. See the How to Simulate
Arrhythmias section.
4. When the waveform group value is highlighted, push .
To change the heart rate:
1. Push or to move the highlight in the display to the Heart Rate value.
2. Push .
3. Push or to adjust the heart rate. Each push of a key moves the heart rate one
beat in the direction of the key pushed. The ProSim 8 simulates a heart rate of 10 to
360 beats/minute (BPM) set in 1 BPM intervals. The ProSim 6 simulates a heart rate
of 30 to 360 beats/minute (BPM) set in 10 BPM intervals.
4. Push to set the heart rate and go back to the ECG screen.
To change the axis:
1. Push or to move the highlight to the Axis value.
2. Push .
3. Push or to highlight Intermediate, Horizontal, or Vertical.
4. Push .
To set the ST Deviation:
1. Push or to move the highlight to the ST Deviation value.
2. Push .