How to simulate invasive blood pressure tests, How to simulate a swan-ganz procedure – Fluke Biomedical ProSim 6 User Manual
Page 43

Vital Signs Simulator
Invasive Blood Pressure Simulation and Tests
31
Note
The artifact values in the list can be in percent (%) or in mmHg. These units
are set by the chamber value.
4. Push to set the artifact value and go back to the IBP screen.
The IBP screen reappears in the display with the new artifact selection.
How to Simulate Invasive Blood Pressure Tests
The Product can simulate pressures that occur during a Swan-Ganz or Cardiac
Catheterization procedure. Figure 20 shows a monitor connected to the two IBP channel
jacks on the Product.
ProSim 8
Patient
Monitor
IBP Cables
glh028.eps
Figure 20. Invasive Blood Pressure Connections
How to Simulate a Swan-Ganz Procedure
To simulate IBP during a Swan-Ganz procedure:
1. Push the Tests softkey in the IBP screen.
2. Push or to highlight Swan-Ganz – Channel 1 or Swan-Ganz –
Channel 2.
3. Push . The details and steps in the procedure shown in Figure 21 are shown in
the display.