ProSim 6/8
Users Manual
40
6
60
70
50
40
ProSim 8
Patient
Monitor
Cardiac Output
Adapter
Blood
Temperature
Connection
Injectate
Temperature
Connection
glh057.eps
Figure 29. Cardiac Output Connections
Setup the DUT for the simulated parameters that follow:
•
Catheter Size: 7 F
•
Injectate Volume: 10 cc
•
Computational Constant: 0.542 or 0.595, based on the injectate temperature.
To set the Product for a cardiac output test:
1. Push .
2. Push , , , or to highlight Cardiac Output and then push to show the
cardiac output parameters in the display. See Figure 30.
glh017.bmp
Figure 30. Cardiac Output Screen
Push the softkey labeled Back to go back to the Special Functions screen,