8 update of non-volatile calibration dates, Update of non-volatile calibration dates -10 – AMETEK M130 User Manual

Page 92

background image

Calibration

Sorensen Ethernet Option

4-10

M130/M131 Programming Manual


12.

Program the supply to store the calibration values in non-volatile memory:

CAL[n]:STORE

Allow 10 seconds for the non-volatile memory to be updated with the
new calibration values.


13.

Program the supply to lock the non-volatile memory for calibration value
protection

CAL[n]:LOCK


14.

The current readback calibration is complete. The unit may be turned OFF or
other calibration procedures may be performed.

4.8 U

PDATE OF

N

ON

-V

OLATILE

C

ALIBRATION

D

ATES

This procedure allows the user to store, in non-volatile memory, the date of the last
calibration (LASTCALDATE) and the date for the next due calibration
(NEXTCALDATE).

1. Program the supply with the date of the most recent calibration:

CAL[n]:MOD:LASTCALDATE

(e.g. CAL1:MOD:LASTCALDATE 07 31 2005)

2. Program the supply with the date of the next calibration (normally 1 year from

last calibration):

CAL[n]:MOD:NEXTCALDATE

3.

Program the supply to unlock the non-volatile memory for calibration value
storage:

CAL[n]:UNLOCK “6867”

4.

Program the supply to store the calibration values in non-volatile memory:

CAL[n]:STORE

Allow 10 seconds for the non-volatile memory to be updated with the
new calibration values.

5.

Program the supply to lock the non-volatile memory for calibration value
protection

CAL[n]:LOCK

6.

The unit may be turned OFF or other calibration procedures may be
performed.