Sorensen Ethernet Option
M130/M131 Programming Manual
4-5
4.4 V
OLTAGE
M
EASUREMENT
/R
EADBACK
C
ALIBRATION
STEP
DESCRIPTION
1.
Program the overvoltage protection to maximum to prevent nuisance trips:
CAL[n]:OUTP:VOLT:PROT:DAC 65535
2.
Program the output current to full scale to prevent Constant-Current
operation:
CAL[n]:OUTP:CURR:DAC 65535
3.
Program the output of the first calibration point to approximately 15% of full
scale voltage by sending the following command string from the computer:
CAL[n]:OUTP:VOLT:DAC 6500
4.
Let the output settle and measure the voltage with the meter.
5.
Enter the actual voltage readback corresponding to the DAC value 6500 of
the first calibration point:
CAL[n]:MEAS:VOLT:POINT 1
6.
Program the output of the second calibration point to approximately 85% of
full scale voltage by sending the following command string from the
computer:
CAL[n]:OUTP:VOLT:DAC 57500
7.
Let the output settle and measure the voltage with the meter.
8.
Enter the actual voltage readback corresponding to the DAC value 57500 of
the second calibration point:
CAL[n]:MEAS:VOLT:POINT 2
9.
Reset the output voltage to 0 volts.
CAL[n]:OUTP:VOLT:DAC 0
10.
Program the M130 to calculate the voltage readback calibration gain and
offset values:
CAL[n]:MEAS:VOLT:CALC
11.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL[n]:UNLOCK “6867”