6 serial poll operation, Serial poll operation -6 – AMETEK M130 User Manual
Page 56

IEEE 488.2 and SCPI Command Operation
Sorensen Ethernet Option
3-6
M130/131 Programming Manual
Error Code Description
-221
Settings conflict
Attempted to set output greater than soft limits or to set soft limits less
than output.
-222
Data out of range
Parameter exceeded range of valid values.
-225
Out of memory
There is not enough memory to perform the requested operation.
-241
Hardware missing
A legal command or query could not be executed because the option is
not installed.
-284
Program currently running
A legal command or query could not be executed because a function is
currently running.
-292
Referenced name does not exist
-293
Referenced name already exists
-316
Checksum error
-330
Self-test failed
A self-test failure has occurred.
-340
Calibration failed
Error during calculation of calibration values occurred.
-350
Queue overflow
The error queue can contain up to 10 entries. If more than 10 error/event
conditions are logged before the SYSTem:ERRor? query, an overflow will
occur; the last queue entry will be overwritten with error -350. When the
queue overflows, the least recent error/events remain in the queue and the
most recent error/events are discarded.
-360
Communication error
Communication to a channel was disrupted.
3.2.6
S
ERIAL
P
OLL
O
PERATION
Performing a serial poll will not modify the Status Byte other than to clear the
RQS (bit 6) for a supply requesting service. Queries affecting the Status
Registers and subsequent serial poll are described below:
• *STB? clears the Status Byte
• *ESR? clears the ESR and bit 5 of the Status Register
• SYSTem:ERRor? clears bit 2 of the Status Register if the queue is empty