Sorensen Ethernet Option
IEEE 488.2 and SCPI Command Operation
M130/M131 Programming Manual
3-13
Command
Description
SCPI
:GAIN
Sets the value of the gain for current
measurements.
N
:OFFSet
Sets the value of the offset for current
measurements.
N
:POINt <1|2> <0+NRf>
Sets the current measurement calibration
point (1 or 2). The actual output current is
measured with an external meter.
N
:VOLTage
N
:ADC?
Returns the integer value of the A/D for the
voltage measurement.
N
:CALCulate
Calculates the value of the gain and offset
for voltage measurements.
N
:GAIN
Sets the value of the gain for voltage
measurements.
N
:OFFSet
Sets the value of the offset for voltage
measurements.
N
:POINt <1|2>
Sets the voltage measurement calibration
point (1 or 2). The actual output voltage is
measured with an external meter.
N
:MODEL
:LASTCALDATE
Sets the date last calibrated.
Order: Month Day Year.
:LASTCALDATE?
Returns the date last calibrated.
:NEXTCALDATE
Sets the due date for next calibration.
Order: Month Day Year
:NEXTCALDATE?
Returns the due date for next calibration.
:OUTPut
N
:CURRent
N
:DAC <0+NR1>
Sets the output of the output current D/A
converter. The value range is 0 – 4095 for
12-bit operations; 0-65535 for 16-bit.
N
:FIVEPOINT <1|2|3|4|5>
<0+NRf>
Sets the five current calibration values.
N
:FIVEPOINT?
Displays the set of five entered calibration
current values.
N
:GAIN
Sets the value of the gain for the output
current.
N
:OFFSet
Sets the value of the offset for the output
current.
N
:VOLTage
N
:DAC <0+NR1>
Sets the output of the output voltage D/A
converter. The value range is 0 - 4095 for
12-bit operations; 0-65535 for 16-bit.
N
:FIVEPOINT <1|2|3|4|5>
<0+NRf>
Sets the five voltage calibration values.
N
:FIVEPOINT?
Displays the set of five entered calibration
voltage values.
N
:GAIN
Sets the value of the gain for the output
voltage.
N