Sorensen Ethernet Option
IEEE 488.2 and SCPI Command Operation
M130/M131 Programming Manual
3-23
Command
Description
SCPI
:PROTection
C
:CONDition?
Returns the integer value of the Protection
Condition Register. Used to read the status
of the power hardware. See section 3.2.3 for
a detailed table of the various bits that make
up this register.
C
:ENABle <0+NR1>
Sets the enable mask of the Protection
Event Register, which allows true conditions
to be reported in the summary bit of the
Protection Condition Register.
C
:EVENt?
Returns the integer value of the Protection
Event Register.
C
:SELEct <0+NR1>
This command provides a means for
selecting which fault bits from the protection
event register (also called the fault register
and readable using the STAT:PROT:EVEN?
command) are able to set the protection
event flag bit in the SCPI status byte
(readable using the *STB? command). It
defaults to value 255 at power-on time, and
never changes unless intentionally
programmed to a new value. Available in the
DLM600 version of the M130 GPIB or
Ethernet option with firmware version 1.05 or
later. Not available in the DCS.
N
:SELEct?
Queries the last selection value
programmed.
N
:QUEStionable
C
:CONDition?
Returns the integer value of the
Questionable Condition Register. The query
is supported but will always return “0”
indicating operational condition.
C
:ENABle <0+NR1>
Sets the enable mask of the Questionable
Event Register allowing true conditions to be
reported in the summary bit of the
Questionable Condition Register. Values are
written and queried but have no effect on the
Questionable Condition Register.
C
:EVENt?
Returns the integer value of the
Questionable Event Register. This query is
supported but always returns a value of “0”,
indicating operational condition.
C