Sorensen Ethernet Option

Calibration

M130/M131 Programming Manual

4-3

4.3 V

OLTAGE

P

ROGRAMMING

C

ALIBRATION

The following procedures are for five-point calibration:

STEP

DESCRIPTION

1.

Issue a *CLS[n] command.

2.

Issue a *RST[n] command.

3.

Program the overvoltage protection to maximum to prevent nuisance trips:

CAL[n]:OUTP:VOLT:PROT:DAC 65535

4.

Program the output current to full scale to prevent Constant-Current
operation:

CAL[n]:OUTP:CURR:DAC 65535

5.

Program the output of the first calibration point by sending the following
command string from the computer:

CAL[n]:OUTP:VOLT:DAC 3275

6.

Let the output settle and measure the voltage with a high precision voltmeter;
this is value 1.

7.

Enter the actual output voltage value of the first calibration point:

CAL[n]:OUTP:VOLT:FIVEPOINT 1

8.

Program the output of the second calibration point by sending the following
command string from the computer:

CAL[n]:OUTP:VOLT:DAC 19000

9.

Let the output settle and measure the voltage with a high precision voltmeter;
this is value 2.

10.

Enter the actual output voltage value of the second calibration point:

CAL[n]:OUTP:VOLT:FIVEPOINT 2

11.

Program the output of the third calibration point by sending the following
command string from the computer:

CAL[n]:OUTP:VOLT:DAC 32000

12.

Let the output settle and measure the voltage with a high precision voltmeter;
this is value 3.