3 voltage programming calibration, Voltage programming calibration -3 – AMETEK M130 User Manual
Page 85

Sorensen Ethernet Option
M130/M131 Programming Manual
4-3
4.3 V
OLTAGE
P
ROGRAMMING
C
ALIBRATION
The following procedures are for five-point calibration:
STEP
DESCRIPTION
1.
Issue a *CLS[n] command.
2.
Issue a *RST[n] command.
3.
Program the overvoltage protection to maximum to prevent nuisance trips:
CAL[n]:OUTP:VOLT:PROT:DAC 65535
4.
Program the output current to full scale to prevent Constant-Current
operation:
CAL[n]:OUTP:CURR:DAC 65535
5.
Program the output of the first calibration point by sending the following
command string from the computer:
CAL[n]:OUTP:VOLT:DAC 3275
6.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 1.
7.
Enter the actual output voltage value of the first calibration point:
CAL[n]:OUTP:VOLT:FIVEPOINT 1
8.
Program the output of the second calibration point by sending the following
command string from the computer:
CAL[n]:OUTP:VOLT:DAC 19000
9.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 2.
10.
Enter the actual output voltage value of the second calibration point:
CAL[n]:OUTP:VOLT:FIVEPOINT 2
11.
Program the output of the third calibration point by sending the following
command string from the computer:
CAL[n]:OUTP:VOLT:DAC 32000
12.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 3.