3 ieee-488.2 and scpi conformance information, 1 parameter definitions, 2 units – AMETEK M130 User Manual
Page 57: Ieee-488.2, Scpi conformance information -7, Parameter definitions -7, Units -7, Table 3-5 parameter definitions -7, Table 3-6 m130 units -7, Scpi
Sorensen Ethernet Option
IEEE 488.2 and SCPI Command Operation
M130/M131 Programming Manual
3-7
3.3 IEEE-488.2
AND
SCPI
C
ONFORMANCE
I
NFORMATION
The supply conforms to most of the specifications for devices as defined in IEEE-
488.2 and SCPI Version 1995.0. Confirmed Commands are those that are
approved commands in the SCPI 1995 Specification, Volume 2: Command
Reference. Confirmed Commands are denoted by a “C” in the “SCPI” column.
Any commands that are not Confirmed Commands are labeled as Not Approved
denoted by an “N” (e.g., Section 3.5.2).
3.3.1
P
ARAMETER
D
EFINITIONS
Table 3-5 Parameter Definitions
PARAMETER DEFINITIONS
Type
Valid Arguments
“ON” or 1. “OFF” or 0.
The data format
Zero, positive and negative integer numeric values are valid data.
<0+NR1>
Zero and positive integer numeric values.
<-NR1>
Negative integer numeric values.
The data format
Representation. Zero, positive and negative floating point numeric
values are some examples of valid data.
<0+NRf>
Zero and positive floating point numeric values.
<-NRf>
Negative floating point numeric values.
Characters enclosed by single or double quotes.
3.3.2
U
NITS
The Series M130 will accept the following units as suffixes to numeric values:
Table 3-6 M130 Units
UNITS
Type of Unit
Valid Suffix
Voltage
“VOLTS” or “volts”, “V” or “v”, “MV” or “mv” or “mV”
Current
“AMPS” or “amps”, “A” or “a”, “MA” or “ma” or “mA”
Time
“SEC” or “sec”, “S” or “s”, “MS” or “ms”, “MIN” or “min”
Frequency
“HZ” or “hz”
The default units are VOLTS, AMPS, SEC, and HZ. For example, “SOUR:VOLT
1” programs 1 volt. To program in units of millivolts, type “SOUR:VOLT 1mV”.